TECHNICAL SPECIFICATIONS ALM-155 - Lab-Wissen

Search our Site
Go to content

Main menu:

TECHNICAL SPECIFICATIONS ALM-155

TECHNICAL SPECIFICATIONS:
Measurement Range:
Alcohol content: 0.00 to 100.00 vol%
Density: 0.69937 to 1.24887 g/cm3
Specific Gravity: 0.70000 to 1.25000
Repeatability:
Alcohol content SD: 0.05 vol%
Density SD: 0.00005 g/cm3
Specific Gravity SD: 0.00005
Accuracy:
Alcohol content: 0.01 vol%
Density: 0.00001 g/cm3
Specific Gravity: 0.00001
Temperature: 0.01°C
Measurement Temperature:
20°C (Fixed)
Alcohol Table:
Selectable either OIML or AOAC
Minimum Sample Required:
8 mL Automatic through pump (for 10 seconds of sampling time)
Measurement Time In automatic operation:
2 to 4 minutes (with peristaltic pump)
Display:
LCD (with a backlight)
Displayed contents:
Density; Specific Gravity; Alcohol Content; Temperature
Sampling:
(1) Auto Sampling by Peristaltic Pump
(2) Manual by optional syringe
Auto Start function:
Number of repeat time: 2-100
Number of Measurement Result:
100
Density-Concentration Conversion:
Alcohol content
Fixed (Stored as Standard)
Memory for Factor:
One Memory Saved
Factor Calibration Possible Only with Pure Water
PC Software:
SOFT-CAP (Data Acquisition Software)
Interfaces:
RS-232: Dot Matrix Printer (COM1), PC (COM2)
USB USB flash drive, keyboard
Data Format:
CSV data saving to USB
Ambient Conditions:
Temperature: 5°C to 35ºC
Humidity: Below 85%RH (no condensation)
Power Supply:
100 to 240 VAC±10%, 50/60Hz
Power Consumption:
~30W
Dimension:
Width: 270 mm; Depth: 402 mm; Height: 163 mm
Weight:
~10 kg
Conformity:
standard CE marking EMC: EN61326-1 LVD: EN61010-1; RoHS2 Directive
 
Back to content | Back to main menu